10.6084/m9.figshare.20291241.v1Magnetic resonance imaging (MRI) in patients with implanted deep brain stimulation (DBS) systems is subject to strict guidelines in order to ensure patient safety. Criteria include limits on the number of implanted leads. Here, we describe the case of a 29-year-old patient with generalized dystonia implanted with 4 DBS electrodes and 2 implantable pulse generators, who had an off-label spinal MRI without regard for manufacturer guidance yet suffered no adverse effects. This suggests that manufacturer guidelines might be overly restrictive with regards to limits on implanted DBS hardware. Further research in this area is needed to widen access to this fundamental imaging modality for patients with DBS.